CardioSense — ECG Insight Lab: Reproducible Machine Learning and Signal Analysis for ECG Data

Academic Research & Graduate Portfolio Prototype for Physiological Signal Processing and Machine Learning
A transparent, leak-free framework bridging historical electrocardiographic signals, decoupled predictive models, and reproducible biomedical engineering.

CI Python Version License: MIT Tests Design System Model Integrity


1. Project Overview & Research Scope

ECG Insight Lab (CardioSense) is an open-source, software-only research platform developed to investigate reproducible signal conditioning, deterministic feature engineering, and decoupled machine learning models across public electrocardiogram benchmarks.

Active Core Workflows

The platform is architected around three active, mathematically decoupled workflows:

  1. Workflow 1: 12-Lead ECG Multi-Label Classification (PTB-XL v1.0.3)
    Multi-label diagnostic classification across 5 SCP-ECG superclasses (NORM, MI, STTC, CD, HYP) using 148 deterministic waveform features and validation-calibrated decision thresholds on official patient-stratified folds.
  2. Workflow 2: Single-Lead Beat Analysis (MIT-BIH Arrhythmia Database v1.0.0)
    Autonomous Pan-Tompkins QRS peak detection, morphological windowing, and 4-class beat classification (N, S, V, Q) evaluated across 21 held-out patient records under strict patient-disjoint partitioning.
  3. Workflow 3: AD8232 Hardware Prototype — Offline Simulation & Acquisition
    Self-contained benchtop analog front-end prototyping (AD8232 + Arduino Uno) with line-oriented UART streaming, offline stream simulation, causal digital filtering, and safety guardrails strictly enforcing electrical safety standards (IEC 60601-1).

Archived Research Module: Optical Paper Digitizer

Note

Archival Status: The Optical Paper Digitizer module has been relocated from the primary navigation and default user experience into an isolated, collapsible archival section. While the classical morphological trace extractor achieves $r = 0.6932$ on synthetic benchmarks, synthetic evaluations do not establish generalization to complex, wrinkled, or low-contrast clinical paper ECGs. The implementation, tests, and synthetic fixtures remain preserved for transparency and ongoing investigation.


2. Research Documentation Suite

For comprehensive evaluation by professors, admissions committees, and technical reviewers, see the dedicated research documentation:


3. Visual Figures & Demonstration Assets

All figures embedded below are genuinely produced from actual project data and reproducible code:

Calibrated 12-Lead Digital ECG Waveform from PTB-XL (Workflow 1)

Calibrated 12-Lead ECG Sample Figure 1: Authentic 10-second 12-lead digital ECG recording (00001_lr) from PTB-XL v1.0.3 sampled at 100 Hz, with physical axes calibrated in seconds and millivolts.


Held-Out Benchmark Performance & Paired Comparison (PTB-XL Fold 10, $N=2,198$)

Model Performance on Fold 10 Paired Patient-Cluster Bootstrap Distribution
Model Performance Metrics Paired Bootstrap Distribution
Figure 2: Full-fold model F1 & PR-AUC vs. pilot model across 5 diagnostic superclasses. Figure 3: Paired difference in Macro F1 ($\Delta = +0.1510$, $95%$ CI $[0.1343, 0.1681]$, $B=2,000$).

Hardware Prototype Benchtop Setup & Real-Time Serial Stream Evidence (Workflow 3)

Benchtop verification of the AD8232 analog front-end and real-time UART serial data transmission:

AD8232 to Arduino Uno Setup Live ADC Waveform in Serial Plotter AD8232 & Microcontroller Wiring Schematic
AD8232 Arduino Uno Setup Live ADC Stream Plotter AD8232 Pinout Wiring Schematic
Figure 4: Physical benchtop prototype — AD8232 single-lead ECG front-end wired to Arduino Uno with active power & lead indicators. Figure 5: Real-time serial acquisition on COM9 (9600 baud) displaying analog P-QRS-T complexes. Figure 6: Microcontroller interface wiring schematic (3.3V, GND, Output ADC, and LO+/LO- leads-off detection).

Interactive Application Video Demonstrations

The complete interactive research interface can be observed in real time across the two core platform modes:

1. Practical Lab & Live Experiment Walkthrough (lab.mp4)

Demonstrates interactive 12-lead signal exploration, zero-phase filtering and Bode response, diagnostic superclass inference, confusion matrices, single-lead MIT-BIH Holter beat classification, and LLM clinical explanations.

Practical Lab & Live Experiment Walkthrough


2. Guided Educational Demo Walkthrough (demo_guide.mp4)

Demonstrates the 9-step pedagogical walkthrough covering cardiac electrophysiology, data leakage prevention, signal conditioning, decoupled machine learning architectures, and calibrated diagnostic decision cutoffs.

Guided Educational Demo Walkthrough


4. Core Architectural Principles & System Design

flowchart TD
    subgraph WF_1["Workflow 1: 12-Lead ECG Classification (PTB-XL)"]
        B1["PTB-XL v1.0.3 Dataset
(21,799 12-Lead Records, PhysioNet WFDB)"] --> B2["Patient-Clustered Folds
(Folds 1-8: Train | Fold 9: Val | Fold 10: Test)"]
        B2 --> B3["Feature Extraction Engine
(148 Deterministic Signal Features)"]
        B3 --> B4["Multi-Label Diagnostic Classifier
(One-vs-Rest Logistic Regression + Validated Cutoffs)"]
        B4 --> B5["Interactive Explorer & Held-Out Benchmark
(Calibrated mV/s Viewer, Clustered Bootstrap CIs)"]
    end

    subgraph DECOUPLING["Strict Architectural Boundaries (Decoupling)"]
        style DECOUPLING fill:#fff2f2,stroke:#d9534f,stroke-width:2px,stroke-dasharray: 5 5
        BOUNDARY["GATED DECISION BOUNDARY
No Paper-to-Digital Classification Bridge
No Single-Lead Heart Attack / MI Diagnosis
No Live Hardware Sensor Claims"]
    end

    subgraph WF_2["Workflow 2: Single-Lead Beat Analysis (MIT-BIH)"]
        C1["MIT-BIH Arrhythmia Database v1.0.0
(48 Records, 47 Patients, 360 Hz)"] --> C2["Patient-Disjoint Splits
(16 Train | 7 Val | 21 Held-Out Test)
Records 201 & 202 strictly co-located"]
        C2 --> C3["Autonomous Signal-Based R-Peak Detector
(Pan-Tompkins Filter — Zero Annotations at Inference)"]
        C3 --> C4["Feature Extraction & Classifier
(47 Descriptors + Balanced Random Forest)"]
        C4 --> C5["Recorded Playback Simulation in Streamlit
(RECORDED DATA REPLAY — NOT LIVE HARDWARE)"]
    end

    subgraph WF_3["Workflow 3: AD8232 Hardware Prototype & Simulation"]
        H1["AD8232 Analog Front-End
(3-Lead Differential Instrumentation Amp)"] --> H2["Arduino Uno ADC Streaming
(115,200 Baud UART, 360 Hz Scheduled Sampling)"]
        H2 --> H3["Offline Transport Simulator & Parser
(Frame Validation, Packet Verification)"]
        H3 --> H4["Causal Digital Conditioning
(0.5-40 Hz Bandpass, 50 Hz Notch)"]
        H4 --> H5["Streamlit Prototype UI
(Persistent Simulation Notice & IEC 60601-1 Guardrails)"]
    end

    subgraph ARCHIVED["Archived Research Module: Optical Paper Digitizer"]
        style ARCHIVED fill:#f9f9f9,stroke:#999999,stroke-width:1px,stroke-dasharray: 3 3
        A1["Input: Paper ECG Image
(Synthetic Sample Fixture)"] --> A2["Image Validation Gate
(Dimension, Aspect Ratio, Decompression Guards)"]
        A2 --> A3["12-Lead Bounding Box Segmentation
(Standard 3x4 Clinical Layout)"]
        A3 --> A4["Morphological Centroid Digitizer
(Classical CPU Baseline)"]
        A4 --> A5["Signal Visualization & Data Export
(Relative Units, Unrecovered Calibration)"]
    end

    ARCHIVED -.->|UNVALIDATED ON REAL SCANS - GATED| BOUNDARY
    BOUNDARY -.->|PROHIBITED| WF_1
    BOUNDARY -.->|PROHIBITED| WF_2
    WF_3 -.->|SIMULATION ONLY - UNCALIBRATED ADC GATED| BOUNDARY
Loading

Architectural Tenets

  1. Strict Subsystem Decoupling:
    • Workflow 1 (12-Lead Waveform ML): Digital waveform feature extraction and diagnostic classification across 5 SCP-ECG superclasses (NORM, MI, STTC, CD, HYP) trained and evaluated on official patient-stratified partitions.
    • Workflow 2 (Single-Lead Beat Analysis): Single-channel beat-type classification using a project-defined 4-class beat grouping (N, S, V, Q, with ventricular fusion beats F merged into Q) on MIT-BIH Holter data, autonomous Pan-Tompkins QRS detection, and simulated recorded playback.
    • Workflow 3 (Hardware Prototype & Offline Simulation): Self-contained firmware, schematics, host protocol, offline simulator, and dedicated Streamlit simulation UI (Hardware Prototype — Offline AD8232 Simulation) under hardware/. Operates completely offline with simulated 10-bit ADC transport, protocol serialization/parsing, simulator inverse conversion, visible per-beat predictions, and post-hoc reference evaluation. Raw ADC counts are never fed into ML classifiers.
    • Archived Module (Optical Paper Digitizer): Morphological trace extraction preserved in an isolated collapsible UI section and documented under historical benchmarks.
  2. Gated Integration Boundary: Optical digitizer outputs and uncalibrated hardware ADC counts are not routed into diagnostic classifiers, and single-lead signals are never used to predict acute coronary syndrome (MI/STTC/CD/HYP).
  3. Cryptographic Model Trust: Model serialization uses SHA-256 sidecar manifests (waveform_lr_model.joblib.manifest.json, single_lead_rf_model.joblib.manifest.json). Untrusted legacy pickles are strictly quarantined in legacy/ and rejected by code policy.
  4. Zero-Persistence Privacy: Uploaded images and signals are processed entirely in volatile RAM buffers (BytesIO). No files, logs, or patient identifiers are ever written to disk or transmitted across networks.
  5. Quiet Academic Editorial Interface: A distraction-free visual environment designed according to research editorial standards (warm ivory #F7F6F2, deep charcoal #252722, and forest green #455D4A).

5. Quantitative Benchmark Results

12-Lead Waveform ML Benchmark (Workflow 1: PTB-XL Fold 10, $N=2,198$)

Evaluated across all 2,198 records ($1,904$ distinct patients) in the held-out PTB-XL Fold 10 test partition:

Diagnostic Superclass Positive Support Decision Threshold Precision Recall F1-Score PR-AUC
NORM (Normal ECG) 963 0.50 0.6995 0.8484 0.7668 0.8128
MI (Myocardial Infarction) 550 0.50 0.4708 0.6891 0.5594 0.5468
STTC (ST/T Changes) 521 0.55 0.5173 0.7159 0.6006 0.5856
CD (Conduction Disturbance) 496 0.55 0.5581 0.6492 0.6002 0.6671
HYP (Hypertrophy) 262 0.65 0.4520 0.6107 0.5195 0.5338
Macro Average 2,198 — 0.5633 0.6667 0.6093 0.6292
Micro Average 2,198 — 0.5828 0.7077 0.6392 —

Paired Model Comparison & Statistical Methodology

A paired patient-cluster bootstrap analysis ($B = 2,000$ replicates, random seed 42) on the Fold 10 test cohort ($N=2,198$):

  • Pilot Model Macro F1: $0.4583$ vs. Full-Fold Model Macro F1: $0.6093$ ($\Delta = +0.1510$).
  • 95% Paired Patient-Clustered Bootstrap CI for Difference: $[+0.1343, +0.1681]$ ($ ext{SE} = 0.0087$).

Single-Lead Beat ML Benchmark (Workflow 2: MIT-BIH Held-Out Test, 21 Patients)

Note

Taxonomy Disclosure: Project-defined 4-class beat grouping: N, S, V, Q, with ventricular fusion beats (F) merged into Q. This is not a standard five-class AAMI EC57 implementation.
In evaluated non-paced records, 98.2% of Class Q beats are ventricular fusion beats (F) rather than unclassifiable beats. Evaluated across 21 disjoint held-out patient recordings from the MIT-BIH Arrhythmia Database v1.0.0.

1. Primary Benchmark: Matched-Detection Conditional Beat-Classification Metrics (N = 47,034)

Evaluates beat classification on detected R-peaks proposed by the automated Pan-Tompkins detector, matched to reference annotations within $\pm 50, ext{ms}$ tolerance ($\pm 18$ samples @ 360 Hz), and retained after feature-window boundary checks:

  • Signal-Based Detector Performance:
    • Total Reference Beats: 47,576
    • Total Detected Peaks: 47,163
    • True Positives (TP): 47,048 | False Positives (FP): 115 | False Negatives (FN): 528
    • Detector Sensitivity: 98.89% | Detector PPV: 99.76% | Detector F1-Score: 99.32%
  • Evaluated Detections (Feature-Valid): 47,034 beats (47,048 matched detections minus 14 boundary-truncated beats)
  • Primary Accuracy Metrics:
    • Macro F1-Score: 0.5274
    • Weighted F1-Score: 0.9011
Beat Class Grouping True Support Precision Recall F1-Score Grouping Description
N 41,883 0.9592 0.9218 0.9401 Normal / Bundle Branch Block Beats (N, L, R, e, j)
S 1,757 0.3555 0.1793 0.2384 Supraventricular Ectopic Beats (A, a, J, S)
V 3,018 0.7781 0.9298 0.8472 Ventricular Ectopic Beats (V, E)
Q 376 0.0489 0.2979 0.0840 Ventricular Fusion (F) & Unclassifiable (Q)
Macro Average 47,034 0.5354 0.5822 0.5274 Unweighted mean across all 4 classes
Weighted Average 47,034 0.9178 0.8896 0.9011 Prevalence-weighted mean

Note

Benchmark Interpretation Note: Matched-detection conditional beat-classification metrics — classification F1 is calculated only for detector peaks matched to a reference annotation and retained after feature-window boundary checks. Detector false positives (115 peaks) and missed reference beats (542 beats, including boundary-dropped matched beats) are reported separately in detector/count-reconciliation metrics and are not included as errors in this classifier confusion matrix.

2. Secondary Benchmark: Annotation-Centred (Oracle) Classifier Reference ($N = 47,561$)

Isolates classifier feature discrimination assuming ideal QRS localization (centred on expert ground-truth annotation indices; 47,561 valid windows after dropping 15 boundary reference annotations):

Beat Class Grouping Reference Support Precision Recall F1-Score Baseline F1
N (Normal / BBB Beat) 42,184 0.9595 0.9247 0.9418 0.9401
S (Supraventricular Ectopic / PAC) 1,782 0.4035 0.2043 0.2712 0.0000
V (Ventricular Ectopic / PVC) 3,201 0.7912 0.9222 0.8517 0.0000
Q (Ventricular Fusion & Other) 394 0.0549 0.3173 0.0936 0.0000
Macro Average 47,561 0.5523 0.5921 0.5396 0.2350
Weighted Average 47,561 0.9192 0.9002 0.9036 0.8338

3. Controlled Research Experiment: Class Imbalance Mitigation

A controlled experiment was conducted evaluating 4 candidate strategies on patient-disjoint splits to tackle minority class degradation without test peeking (see docs/reports/single_lead_imbalance_experiment.md):

  • Candidate 3 (Minority-Boosted Cost-Sensitive RF): Boosted rare Class S recall from $20.43%$ to $27.10%$ ($+6.68%$) and F1 from $0.2712$ to $0.3245$, with Class V F1 reaching $0.8889$.
  • Admissions Integrity Decision: The balanced baseline model was preserved as production to prevent post-hoc test peeking, and Candidate 3 is thoroughly documented as a research candidate.

Archived Module Benchmark: Optical Digitizer (Synthetic Reference)

Evaluated strictly on the deterministic synthetic 12-lead benchmark fixture (sample_data/synthetic_12lead.png vs. sample_data/synthetic_12lead_gt.csv):

Evaluation Metric Measured Value Acceptance Threshold Status
12-Lead Detection Rate 12 / 12 (100%) $\ge 10 / 12$ leads PASS
Mean Pearson Correlation ($r$) 0.6932 $> 0.60$ PASS
Mean Absolute Error (MAE) 0.0725 (relative units) $< 0.15$ PASS
Inference Latency (CPU) $�pprox 210, ext{ms}$ $< 2,000, ext{ms}$ PASS
Calibration Status UNRECOVERED Honest designation PASS

6. Educational Teaching Labs (Jupyter Notebooks)

The repository provides four interactive, pre-executed computational notebooks under notebooks/:

Notebook Focus & Educational Content Execution Guide
00_system_architecture_and_diagnostics.ipynb Historical evolution, baseline audit trail, and decoupled architecture diagram. Open Guide
01_ecg_image_digitization.ipynb Input validation, classical morphological digitizer execution, ground-truth accuracy. Open Guide
02_ptbxl_data_and_features.ipynb PTB-XL composition, zero-leakage split verification, 148 signal feature extraction. Open Guide
03_model_predictions_and_evaluation.ipynb Verified model loading, Fold 10 inference, full benchmark table, paired bootstrap. Open Guide

7. Installation & Quick Start

Prerequisites

  • Operating System: Linux, macOS, or Windows
  • Python: 3.10, 3.12, or 3.14 (tested and verified on Python 3.14.4)
  • Memory: $\ge 1.0, ext{GiB}$ RAM available

Step 1: Environment Setup

# Clone the repository and navigate to the code directory
git clone https://github.com/omkar703/ECG_Insight_Lab_CardioSense.git
cd ECG_Insight_Lab_CardioSense/code

# Create and activate a clean virtual environment
python3 -m venv .venv
source .venv/bin/activate  # On Windows: .venv\Scripts\activate

# Install production dependencies
pip install --upgrade pip
pip install -r requirements.txt

Step 2: Run Automated Quality Gates

Execute the complete test suite:

pytest -v

Expected Result: 196 passed tests across 15 test suites (tests/ and hardware/tests/).

Step 3: Run System Diagnostics via CLI

Verify system environment, dataset status, and cryptographic model integrity:

python -m ecg_lab.waveform_ml.cli diagnostics

Expected Output: All 7 health checks report [OK] and system reports OPERATIONAL.

Step 4: Launch the Full-Stack Web Application

streamlit run app.py

Open your web browser at http://localhost:8501. The web platform provides two primary modes:

  1. Research Guide: 9-step pedagogical walkthrough of cardiac electrophysiology, noise filtration, Pan-Tompkins QRS cascade, ML architectures, and audited benchmark results.
  2. Practical Lab: Direct backend-connected interactive workspace for 12-lead PTB-XL classification, single-lead MIT-BIH beat analysis, and offline AD8232 simulation.

8. Command Line Interface (CLI) Guide

The unified CLI provides reproducible commands for the entire data and ML pipeline:

# 1. Run system diagnostics and health audit
python -m ecg_lab.waveform_ml.cli diagnostics

# 2. Validate dataset metadata and verify zero patient leakage
python -m ecg_lab.waveform_ml.cli validate --verify-checksums

# 3. Extract 148 deterministic signal features (with caching)
python -m ecg_lab.waveform_ml.cli extract-features

# 4. Train model pipeline (Folds 1-8 train, Fold 9 validation threshold tuning)
python -m ecg_lab.waveform_ml.cli train

# 5. Evaluate frozen model on held-out test fold (Fold 10)
python -m ecg_lab.waveform_ml.cli evaluate

# 6. Audit paired model comparison on Fold 10
python -m ecg_lab.waveform_ml.cli audit-paired-comparison

9. Repository Structure & File Schema

A complete classification of tracked source code, gitignored raw datasets, and quarantined legacy artifacts is detailed in docs/project_file_schema.md:

code/
├── .github/workflows/ci.yml          # GitHub Actions continuous integration workflow
├── app.py                            # Streamlit research web application (Quiet Academic UI)
├── requirements.txt                  # Production dependencies
├── requirements-notebooks.txt        # Optional notebook dependencies
├── ecg_lab/                          # Core modular package (waveform_ml, single_lead_ml, digitization, explanations)
├── notebooks/                        # 4 computational research notebooks & guide
├── tests/                            # 150 unit & integration tests across 15 suites
├── hardware/                         # AD8232 + Arduino Uno prototype, simulation & 46 tests
├── sample_data/                      # Synthetic 12-lead benchmarks & MIT-BIH sample records
├── docs/                             # Technical reports, admissions documentation & audits
│   ├── PORTFOLIO_REPORT.md           # Master admissions portfolio summary report
│   ├── ARCHITECTURE.md               # Complete architecture & data boundary specifications
│   ├── RESEARCH_METHODS.md           # Scientific methodology & signal processing math
│   ├── RESULTS_AND_LIMITATIONS.md    # Audited results, confusion matrices & honest caveats
│   ├── assets/                       # Publication-grade authentic scientific figures & videos
│   ├── reports/                      # 17 formal engineering milestone & experiment reports
│   └── project_file_schema.md        # Full repository taxonomy
├── artifacts/                        # Trained models (12-Lead LR, Single-Lead RF), manifests & experiments
├── legacy/                           # Quarantined legacy pickle models (loading blocked)
└── data/                             # [GITIGNORED] PTB-XL and MIT-BIH raw datasets

10. Academic Provenance, Licensing & Citations

Open-Source License

The application codebase and documentation are licensed under the MIT License.

Data & Literature Citations

  • PTB-XL ECG Dataset:
    Wagner, P., Strodthoff, N., Bousseljot, R. D., Kreiseler, D., Lunze, F. I., Samek, W., & Schaeffter, T. (2020). PTB-XL, a large publicly available electrocardiography dataset. Scientific Data, 7(1), 154. https://doi.org/10.1038/s41597-020-0495-6. Distributed under CC BY 4.0.
  • PhysioNet Repository:
    Goldberger, A., et al. (2000). PhysioBank, PhysioToolkit, and PhysioNet: Components of a new research resource for complex physiologic signals. Circulation, 101(23), e215–e220. https://doi.org/10.1161/01.CIR.101.23.e215.
  • MIT-BIH Arrhythmia Database:
    Moody, G. B., & Mark, R. G. (2001). The impact of the MIT-BIH Arrhythmia Database. IEEE Engineering in Medicine and Biology Magazine, 20(3), 45–50. https://doi.org/10.1109/51.932724.
  • Open-ECG Literature:
    Akershus University Hospital AI in Medicine (Ahus-AIM). Electrocardiogram-Digitization. Evaluated under CC BY-SA 4.0.

For detailed dependency licenses, see docs/data_sources_and_licenses.md.

About

Research platform for 12-lead paper ECG digitization, calibrated PTB-XL multi-label classification, and MIT-BIH single-lead arrhythmia analysis with Groq clinical explanations.

Topics

Resources

Stars

1 star

Watchers

0 watching

Forks

Releases

Packages

Contributors

Languages